Specifications
Conductivity |
Measuring range
Resolution Accuracy Temp compensation range Electrode constant |
Conductivity: (0.00 ~ 20.00) μS/cm (20.0 ~ 200.0) μS/cm (200 ~ 2000) μS/cm (2.00 ~ 20.00) mS/cm (20.0 ~ 200.0) mS/cm Resistivity: (0 ~ 100) MΩ·cm TDS: (0 ~ 100) g/L Salinity: (0 ~ 100) ppt 0.01/0.1/1μS/cm 0.01/0.1mS/cm Electrode: ±1.0% FS, Instrument: ±1.50%FS (0 ~ 100)? (Automatic) 0.1/1/10 cm-1 |
Temperature |
Measuring range Resolution Accuracy |
0 ~ 100? 0.1? 5 ~ 60?: ±0.5?, Others: ±1.0? |
Other parameters |
Data storage Power Communication connectors Dimensions & Weight |
200 sets DC 5V/1A Micro USB and standard USB 190*130*18mm/0.31KG |
Working conditions |
Ambient temperature Ambient humidity IP grade |
(5 ~ 35)? ≤85% IP54 |
Standard Configurations
1. L-511 conductivity meter 1 unit
2. Conductivity electrode 1 pc
3. Electrode holder 1 pc
4. 5V/1A power adapter 1 pc
5. Instruction manual 1 copy